Medical Safety Isolation Transformers

OVERVIEW

The IT systems shoul be used in the operation rooms of the hospitals according to TS IEC 61364-7-710. These systems protect the patients and users as well as appliances used in the hospitals. These medical appliances require high class isolation level and well designed according to EN-61558-2-15.

Show more

Our motor drive input reactors comply with European standards and are CE marked Also manufactured under ISO9001: 2008 quality management system

FEATURES

Standart IEC EN 61558-2-15

Power from 3 kVA up to 10 kVA

Working frequency up to 500 Hz

Input Voltage up to 1000 V

Output Voltage up to 1000 V

Protection class IP 00 

Cooling is natural 

Isolation class F – 155C

Isolation test up to 3,5 kV 

Lamination 1,5 W/kg Loss

Ambient tempreture from -15 up to 40

Frequency 50/60 Hz

TECHNICIAL SPECIFICATIONS

Stock No Power a b c h1 h2 Weight(kg)
MED:003 3 KVA 200 260 200 300 340 29
MED:004 4 KVA 250 260 200 350 390 36
MED:005 5 KVA 250 300 220 350 410 42
MED:006 6 KVA 250 320 240 350 410 55
MED:008 8 KVA 300 300 220 250 360 60
MED:010 10 KVA 300 380 300 400 460 75